SEC-MALS
Size-Exclusion Chromatography with Multi-Angle Light Scattering (SEC-MALS)
SEC-MALS couples a size-exclusion chromatography to multi-angle light scattering, a refractive-index (concentration) detector, and UV absorption. The combination yields the absolute molar mass of each eluting species — independent of size-exclusion chromatography calibration or molecular shape — due to scattering intensity being proportional to molar mass times concentration.
Experiment Examples
- Determining whether a purified protein is a monomer, dimer, or higher oligomer in solution, with an exact mass determination.
- Measuring the absolute mass of protein/protein or protein/nucleic acids complexes to determine stoichiometry.
- Quantifying the percentage of soluble aggregate (high-molecular-weight species).
- Measuring the protein and glycan mass fractions of a glycoprotein separately.
- Determining the mass of the protein component of a detergent-solubilized membrane protein, excluding bound detergent.
Sample Requirements
- Standard molar-mass determination: 50 – 200 µg per injection; concentration 1 – 5 mg/mL in 60 µL.
- Complex formation analysis: Sample containing components separately and in complex. Concentrations and volumes as above; amounts comparable for each sample. Ensure the sample is in the final running buffer (same buffer used to equilibrate the column).
- Conjugate analysis (glycoproteins, PEGylated proteins, detergent-bound membrane proteins): Requires slightly more material.
- Membrane proteins: Similar requirements as above. Use detergent beloe its CMC in the running buffer;
- Buffer constraints: Filtered (0.05 µm), degassed, particulate-free running buffer is essential. Salt ≥150 mM is recommended to reduce non-specific column interactions.
Practicalities
- Running buffer should be made fresh, filtered, and provided a day before. The system needs to be equilibrated overnight, at minimum, before measurements.
- The sample has to be filtered with 0.2 µm filter 1.5 mL reaction tube filters. Otherwise bring 65 µL of sample to be filtered on-site.Buffer
- Preparation of Running Buffer:
You will need: a 0.2 µm filter unit (single-use or with exchangeable membrane), two 500 mL bottles, and Milli-Q/ddH₂O.
- Pre-rinse the filter
- Attach the filter unit to a 500 mL bottle.
- Filter 500 mL of Milli-Q/ddH₂O.
- Discard the filtered water and the bottle.
- Rinse the buffer bottle
- Move the filter unit to a new bottle, which will hold your buffer.
- Filter 250 mL of Milli-Q/ddH₂O.
- Remove the filter unit, swirl the water to rinse the bottle walls, and discard the water.
- Repeat steps 2.2–2.3 once more, for two rinses in total.
- Filter the buffer
- Place the filter unit back on the rinsed bottle.
- Filter your buffer.