Nipkow laser photo-manipulations + Temp Controller + CO2 (Tel. 94 787)
Capabilities Confocal_Spinning_Disk.pdf
- Fast Imaging living samples
- Confocal microscopy (2 pinholes' disks, 25 um in diameter for 20x/40x objectives and 50 um in diameter for 63x/100x objectives)
- Widefield Imaging
- Temperature (21°C, 37°C), humidity, CO2 control (top stage incubator, OKO lab), autofocus system
- Fast dual-colour imaging (2 cameras), ratiometric imaging, FRET technique
- Photo-manipulations (-activation, -conversion, -bleaching)
- Laser injury

- Zeiss Axio Observer 7 (inverted)
- CSU W1 automatic Yokogawa spinning disk head (2 disks, 25 and 50 um), 2 camera ports
- Definite Focus 3 system
Objectives
|
Calibration camera bin 1
| Objectif | um / pixel |
| 5x | 1.274 |
| 10x | 0.650 |
| 20x | 0.315 |
| 40x | 0.163 |
| 63x | 0.103 |
| 100x | 0.063 |
Zstack Nyquist calculation
| for laser 488nm | for laser 637nm | |
| 100x | 0.10 um | 0.15 um |
| 63x | 0.11 um | 0.33 um |
| 40x | 0.16 um | 0.47 um |
| 20x | 0.29 um | 0.88 um |
| 10x | 2.5 um | 7.6 um |
| 5x | 9 um | 27 um |
Widefield illumination
pE-800 high-power LED illumination system
Yokogawa CSU
Dichroic:
405/488/561/640 nm
445/515/640
Emission filter well 1 camera straight:
460/50
525/50
535/30
605/52
700/75
position 6-10 Empty
Emission-filter well 2 camera back:
460/50
470/24
525/50
position 4-10 Empty
Camera port:
561 LP
515 LP
Dummy glass
Lasers for confocal imaging and Photo-manipulations
- 405 nm (300 mW)
- 445 nm (100 mW)
- 488 nm (200 mW)
- 515 nm (100 mW)
- 561 nm (150 mW)
- 637 nm (140 mW)
+ pulsed-laser @UV355 nm
+ 2D Visi-FRAP control 2D scanner
Detectors
PRIME BSI Express back-lluminated cameras
Characteristics:
sCMOS camera x 2
Combined with high sensitive back-illumination
95% Quantum Efficiency
6.5 um x 6.5 um Pixel area
95 Frames per second over USB 3.2 Gen 2
Inserts
35 mm Petric-dish holder (magnetic)
Conventional slide holder (magnetic)
Computer
- Windows 11 professional
- Visiview 7.00.3